TROUBLESHOOTING
7-20 Sysmex KX-21 Operator’s Manual -- Revised October 1998
Error Message Meaning of Message
"HGB Error"
*
Help
*
3:Cancel
Clean TD Chamber will be performed.
(It will take approx. 7 minutes.)
[1]
Error occurred during HGB analysis.
Flow Cell may be dirty.
ERR CODE:XXXXX.X.X
1:Exec. Clean
HGB converted value is out of the prescribed
range.
1. Blank value is out of the range.
2. Blank value exceeds sample value.
Cause 1) Dirty HGB flow cell
2) Bubbles mixed in HGB sample line
3) Dirty WBC TD chamber
Action 1) Clean Transducer
Press [1] key to execute the transducer rinse sequence.
Refer to
Chapter 4, Section 5.2: Clean Transducer
.
Analysis after error
occurrence
Ready for analysis (although the next sample could be affected)
NOTE: Press [3] key to return to the Analysis screen. Although the system is
ready for analysis, the error may occur in the next analysis.