TROUBLESHOOTING
Sysmex KX-21 Operator’s Manual -- Revised October 1998 7-21
Error Message Meaning of Message
"Analysis Error"
ERR CODE:XXXXX.X.X
*
Help
*
3:Return
WBC/HGB Analysis Error
WBC (including HGB) analysis error is
continuing.
Cause 1) Faulty lyse (STROMATOLYSER-WH)
Action 1) Replenish Lyse
After replenishing lyse, always execute "4: Replace Reagent" in
the Select Menu, making reference to
Chapter 4, Section 8.1:
Replenish Reagent
. After replenishment and checking, analyze
the control blood and confirm that there is no abnormality.
Analysis after error
occurrence
Ready for analysis (although the next sample could be affected)
NOTE: Press [3] key to return to the Analysis screen. Although the system is
ready for analysis, the error may occur in the next analysis.