TROUBLESHOOTING
7-22 Sysmex KX-21 Operator’s Manual -- Revised October 1998
Error Message Meaning of Message
"Analysis Error"
ERR CODE:XXXXX.X.X
*
Help
*
3:Return
Abnormal detection sensitivity
Sensitivity in the transducer exceeds the
prescribed range.
Cause 1) Faulty diluent (CELLPACK)
Action 1) Replenish Diluent
After replenishing diluent, execute "5: Auto Rinse" in the Select
Menu, and perform background check. In addition, analyze the
control blood to see that there is no abnormality.
Analysis after error
occurrence
Ready for analysis (although the next sample could be affected)
NOTE: • Press [3] key to return to the Analysis screen. Although the system is
ready for analysis, the error may occur in the next analysis.