TROUBLESHOOTING
Sysmex KX-21 Operator’s Manual -- Revised October 1998 7-23
3.7 Memory Errors
Error Message Meaning of Message
"Memory Error"
ERR CODE:XXXXX.X.X
*
Help
*
Memory error occurred.
Turn OFF then ON the power.
Error occurs on the Read Only Memory (ROM)
or Random Access Memory (RAM). The alarm
keeps sounding.
Cause 1) Momentary power failure, sudden noise interference, etc. causes
CPU malfunction.
Action 1) Turn OFF then ON the power again.
Analysis after error
occurrence
Inoperative (The power could be turned OFF.)
Error Message Meaning of Message
"Set Value Error"
ERR CODE:XXXXX.X.X
*
Help
*
Memory error occurred.
Stored data,and
QC data
will be repaired.
Repairing may end incomplete.
2 Deleting files and initializing.
[1]
[2]
1:Repairing 2:Initialize
Errors occur in the set values of the following:
Stored data
Quality Control data
Customer-set values
Cause 1) Momentary power failure, sudden noise interference, etc. causes
analysis value errors.
Action
1) Repair
Press [1] key to execute repairing.
2) Initializing
If execution of "1: Repairing" fails to correct the error, initialize
using [2] key. Initializing deletes all stored data and QC data, and
set values return to initial values.
After completion of initializing, perform all settings referring to
Chapter
10: INSTRUMENT SETUP.
Analysis after error
occurrence
Not Ready until the error is resolved.