TROUBLESHOOTING
7-24 Sysmex KX-21 Operator’s Manual -- Revised October 1998
Error Message Meaning of Message
"Set Value Error"
ERR CODE:XXXXX.X.X
*
Help
*
Memory error occurred.
Turn OFF then ON the power.
Errors occur in the set values of the following:
• Stored data
• Quality Control data
• Customer-set values
• Factory-set values
Cause 1) Momentary power failure or sudden noise interference causes set
value errors.
Action 1) Turn OFF then ON the power again.
If the error persists even after the power is turned OFF then ON,
contact your Sysmex service representative for assistance.
Analysis after error
occurrence
Inoperative (The power could be turned OFF.)
Error Message Meaning of Message
Momentary Power Failure
Momentary power failure occurred, or
[SHUT DOWN]has not been performed
in the last time.
Perform[SHUT DOWN]in prior to
turning OFF the power.
Continue to start up.
1:Continue
[1]
The program had not been normally quitted the
last time the power was turned off.
Cause 1) Momentary power failure, etc. caused the unit to stop temporarily.
2) Shutdown had not been performed the last time the unit was
turned OFF.
Action Press [1] key to start up the unit. Perform SHUTDOWN before turning
OFF the power.
Analysis after error
occurrence
Ready for analysis (This error occurs at starting.)