ADJUSTMENT
8-4 Sysmex KX-21 Operator’s Manual -- Revised October 1998
NOTE: The Status Display screen shows current pressure and vacuum.
SEQ.No. 12
1 2 3 4
SV 1 2 3 4 5 6 7 8 9 0
SENSOR 1 2 3 4 5 6 7
HGB CONVERT 550
VACUUM 240
1 2 3 4 5 6 7 8 9 0
PRESSURE 0.52
WB
Ready
*
Status
*
Vacuum: 250 mmHg
(displaying 240 mmHg)
Pressure: 0.5 kg/cm
(displaying 0.52 kg/cm )
2
2
Change screen with [ ] or [ ]
Figure 8-2-5: Status Display Screen
(6) Adjust the pressure and vacuum by following the procedures given on the next page