ADJUSTMENT
Sysmex KX-21 Operator’s Manual -- Revised October 1998 8-5
2.3 Adjusting Pressure to 0.5 kg/cm
2
(1) Using the procedure of "2.2 Pressure and vacuum Display," display the Status Display
screen.
(2) Loosen the 0.5 kg/cm
2
regulator locking nut on the left side of the unit.
(3) While watching the pressure and vacuum indication on the Status Display screen, turn
the adjusting knob to regulate pressure and vacuum. Turn the adjusting knob
clockwise to increase pressure and vacuum.
Adjustment range: 0.5±0.1 kg/cm
2
Low
High
Locking nut
Adjusting knob
Loosen
Figure 8-2-6: Adjusting the Pressure
CAUTION: • Always adjust pressure and vacuum so as to increase to the
predetermined levels. Should any of them be too high, lower it to
below that level once, then raise it to the set level. Unless adjustment
is properly conducted, accurate adjustment may not be possible.
(4) When adjustment is finished, tighten the locking nut while taking care not to allow the
adjusting knob to rotate.
(5) Press [SELECT] key to return to the Ready status.