INSTRUMENT SETUP
Sysmex KX-21 Operator’s Manual -- Revised October 1998 10-11
(5) Press [ENTER] key. The current setting status is displayed on the Patient Limit
screen.
LL UL
HGB 8.0 17.0 g/dL
RBC 2.50 5.50
×10
6
/µL
WBC 3.0 15.0 ×10
3
/µL
MCV 86.0 110.0 fL
HCT 26.0 50.0 %
MCH 26.0 38.0 pg
PLT 50 400
×10
3
/µL
MCHC 31.0 37.0 g/dL
LYM% 5.0 55.0 %
MXD% 1.0 20.0 %
*
Patient Limit
*
Press [SELECT] to exit.
Figure 10-4-2: Patient Limit Screen
(6) Using [
] or [ ] key, move the cursor to select the desired set parameter.
NOTE: Total 19 setting parameters of Patient Limit are displayed. But not all of
them can be displayed in one screen at a time. Therefore, scroll the
screen using [
] or [ ] key.
LL UL
LYM# 0 0
×10
3
/µL
NEUT% 45.0 95.0 %
MXD% 1.0 20.0 %
NEUT# 0 0
×10
3
/µL
MXD# 0 0 ×10
3
/µL
RDW-SD 37.0 54.0 fL
PDW 9.0 17.0 %
RDW-CV 11.0 16.0 %
MPV 9.0 13.0 fL
P-LCR 13.0 43.0 %
*
Patient Limit
*
Press [SELECT] to exit.
Figure 10-4-3: Scrolling the Screen
(7) Using [ ] or [ ] key, move the cursor to select LL (Lower Limit) or UL (Upper
Limit).
(8) Enter the settings using the numeric keys.