INSTRUMENT SETUP
Sysmex KX-21 Operator’s Manual -- Revised October 1998 10-13
5. QC SETTINGS
This program is used to set QC method and output method of QC data. Two kinds of QC
methods are available with this instrument:
X
control and L-J control. Select either of them
that suits your purpose.
X
Control: Control blood is subjected to two consecutive analyses and the mean of
them is used as the QC data. This method causes little influence on
reproducibility in analysis.
L-J Control: This control uses the data from a single analysis of control blood as QC
data. The control width in the L-J control is prone to influence in
reproducibility in analysis, so that the control width is wider than in the
case of
X
control.
How to Set QC Settings
(1) Press [SELECT] key on the Analysis screen. The Select Menu screen appears.
(2) Using [ ] or [ ] key, move the cursor to select "6: Settings."
(3) Press [ENTER] key. The Setting Menu screen appears.
*Settings*
1:System Setup
2:Date/Time
3:Patient Limit
4:QC Settings
5:Host Settings
6:Printer Settings
7:Print Set Values
Figure 10-5-1: Setting Menu Screen
NOTE: • "5: Host Settin
s" is displa
ed onl
when the optional serial interface is
incorporated.
(4) Using [ ] or [ ] key, move the cursor to select "4: QC Settings."
Sysmex KX-21 Operator’s Manual -- Revised January 2000