EasyManua.ls Logo

SYSMEX KX-21 - Page 267

SYSMEX KX-21
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INSTRUMENT SETUP
10-14 Sysmex KX-21 Operator’s Manual -- Revised October 1998
(5) Press [ENTER] key. The current setting status is displayed on the QC Settings screen.
Data Output Print
*
QC Settings
*
Press [SELECT] to exit.
QC Method
X
Figure 10-5-2: QC Settings Screen
(6) Using [ ] or [ ] key, move the cursor to select the set parameter. QC settings can be
made on two parameters.
(7) Using [ ] or [ ] key, select the content of settings.
The content of settings changes each time you press the key.
(8) Press [ENTER] key. The content of setting is set and the cursor moves to the next
analysis parameter.
NOTE: The settings can also be set when you press [ ] or [ ] key after
changing the content of settings.
1) QC Method Setting
Set the QC method. The QC method can be selected from the following two
kinds: 17
Settings QC Method
X X
Control
L-J L-J Control
Table 10-5-1: QC Method Setting

Related product manuals