EasyManua.ls Logo

SYSMEX KX-21 - Page 275

SYSMEX KX-21
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INSTRUMENT SETUP
10-22 Sysmex KX-21 Operator’s Manual -- Revised October 1998
(6) Using [ ] or [ ] key, move the cursor to select the set parameter. Printer settings can
be made on three parameters.
(7) Using [ ] or [ ] key, select the condition.
The condition changes each time you press the key.
(8) Press [ENTER] key. The content of settings is set and the cursor moves to the next set
parameter.
NOTE: The settings can also be set when you press [ ] or [ ] key after
changing the content of settings.
1) Connect Settings
Set whether to connect the built-in printer or not.
Settings Connection
Use Connect
Not Use Not Connect
Table 10-7-1: Built-in Printer Settings
2) Print Condition Settings
Set the auto print condition for the built-in printer. This can be selected from the
following three kinds:
Settings Auto Print Conditions
All Data Print All Samples
Abnormal Data Print Abnormal Data Only
None Not Auto Print
Table 10-7-2: Print Condition Settings
NOTE: Data printed as abnormal data:
Data in which analysis error occurred
Data in which noise error occurred
Data with flags other than Mark Limits
Data with histogram flag
Data out of the linearity limit
For print conditions of QC data, refer to
Chapter 10, Section 5: QC
Settings.