INSTRUMENT SETUP
10-26 Sysmex KX-21 Operator’s Manual -- Revised October 1998
(4) Using [ ] or [ ] key, move the cursor to select the desired external device.
(5) Using [ ] or [ ] key, select "Use" or "Not Use". "Use" and "Not Use" change over
each time you press the key.
(6) Press [ENTER] key. The content of the settings is set and moves the cursor to the
next set parameter.
NOTE: • The settings can also be set when you press [ ] or [ ] key after
changing the content of settings.
(7) Press [SELECT] key. The Setting Change Confirmation message appears.
Change? Cont. Set Cancel
Figure 10-9-3: Setting Change Confirmation Message
(8) Using [ ] or [ ] key, move the cursor to select "Cont.," "Set," or "Cancel."
[Cont.]: Returns to the Peripheral Settings screen. And the setting operation can
be continued.
[Set]: Updates the settings and returns to the Analysis screen.
[Cancel]: Cancels the changed settings and returns to the Analysis screen.
(9) Press [ENTER] key to execute the selected process.