INSTALLATION
Sysmex KX-21 Operator’s Manual -- Revised October 1998 A-7
Tube Polyurethane 4 mmID × 6 mmOD
Diluent Inlet Nipple
Cubitainer Spout Kit No. 1
CELLPACK
Figure A-5-2: Connecting the CELLPACK
Take care not to spill a reagent on to the instrument. If it spills, wipe it
off immediately using a wet cloth or the like.
If it happens to adhere to your skin, wash it off using plenty of water.
If a reagent happens to enter your eye, wash it off immediately using
plenty of water, and take medical treatment at once.
After unpacking, be sure not to allow dust or dirt to come into the
instrument. Use up CELLPACK within 60 days from the unpacking
date, and STROMATOLYSER-WH within 90 days.
CAUTION: When the diluent (CELLPACK) tube is more than 2 m long, it may be
impossible to have reagent aspirated. Avoid setting CELLPACK at a
level higher than the instrument; otherwise, reagent may flow into the
vacuum line, possibly damaging the instrument.
CAUTION: After connecting the tube, do not pull it by force for reagent
replacement.
CAUTION