DISPLAY AND PROCESSING OF ANALYSIS RESULTS
3-14 Sysmex KX-21 Operator’s Manual -- Revised October 1998
3) Analysis Data
Analysis data of 18 parameters are displayed in a group of four parameters on five
pages:
Stored Data Screen [1/5]: WBC, RBC, HGB, HCT
Stored Data Screen [2/5]: MCV, MCH, MCHC, PLT
Stored Data Screen [3/5]: WBC, LYM%, MXD%, NEUT%
Stored Data Screen [4/5]: WBC, LYM#, MXD#, NEUT#
Stored Data Screen [5/5]: RDW, PDW, MPV, P-LCR
The marks on the left of analysis data indicate the following:
1. ! : The data is out of the linearity limit.
2. + : The data exceeds the upper Mark Limits.
- : The data exceeds the lower Mark Limits.
3. * : The data is low in reliability.
NOTE: • The marks are added with the priority order of 1, 2, and 3.
• The Mark Limits with "+" and "-" can be set by the customer. For
details, refer to
Chapter 10, Section 4: Patient Limit
.
When an analysis error, etc. has caused errors in analysis data, the marks below are
attached to the affected data:
1.
+ + +. +: When data has exceeded the display range.
2. * * *. *: When data could not be calculated due to instrument failure.
3. - - -. -: When data could not be calculated due to data error.
NOTE: • When analysis is made in the pre-diluted mode, particle distribution
analysis parameters are displayed in "- - -. -."
4) Date Analyzed
The date when the analysis data in the underline cursor position was analyzed is
displayed.
5) Manual Discrimination Data
As to data changed through manual discrimination, analysis values are reverse-
displayed.