MAINTENANCE AND SUPPLIES REPLACEMENT
Sysmex KX-21 Operator’s Manual -- Revised October 1998 4-9
5. MONTHLY MAINTENANCE AND PROCEDURE
5.1 Clean Waste Chamber (Rinse Sequence)
When the power is turned on, the message is displayed once a month or for every 2500
samples to instruct the scheduled maintenance.
*
Scheduled Maint.
*
1:Exec.Clean
3:Cancel
Clean the Waste Chamber.
It will take approx. 15 minutes.
Cycle No. after cleaning. 2567
Date last serviced 31/12/97
Figure 4-5-1: Scheduled Maintenance Message Screen
When this message appears, press [1] key to select "1: Exec. Clean," and clean the waste
chamber by the following procedure:
NOTE: • When you press [3] key and select "3: Cancel," the start-up process
continues without cleaning the waste chamber. Until the waste
chamber is cleaned, the above message is displayed at start-up.
• Even when the above message is not displayed, you can execute the
waste chamber cleaning by choosing "7: Maintenance" on the Select
Menu screen, then "1: Clean W. Chamber" on the Maintenance Menu
screen.