MAINTENANCE AND SUPPLIES REPLACEMENT
4-10 Sysmex KX-21 Operator’s Manual -- Revised October 1998
(1) Press [1] key to select "1: Exec. Clean" on the Scheduled Maintenance Message
screen. The Clean W. Chamber screen will display the message instructing the
following operation:
*
Clean W.Chamber
*
Cycle No. after cleaning. 2567
Date last serviced 31/12/97
3:Cancel
Ready
Aspirate CELLCLEAN
It will take approx. 15 minutes.
Figure 4-5-2: Clean W. Chamber Screen
NOTE: • When you press [3] key and select "3: Cancel," the start-up process
continues without cleaning the waste chamber. Until the waste
chamber is cleaned, the above message is displayed at start-up.
(2) Set CELLCLEAN to the sample probe and press the start switch in that status. While
"Aspirating" is being displayed on the screen, keep holding CELLCLEAN in the same
status.
Start Switch
Figure 4-5-3: Pressing the Start Switch
• Since CELLCLEAN is a strong alkaline detergent, take care not to have
it adhere to the skin or clothes. If the skin or clothes should come in
touch with it, flush it away using plenty of water. Otherwise, it can
damage the skin or clothes.
WARNING