V/D
S
ERIES
P
ARAMETERS
8-10 Revision: 2.1
No.
Parameter Description Access
level
Range of values Factory
setting
75 ElectrLevel
(Global 1)
Switching value for water level
electrodes:
999 = Insulation = no water
0 = Short-circuit = water
4 0 – 999 714
or
850 (for
Eco/Basic)
76 WaterTimeAdd
(Global 3)
Run-on time for the water feed pump.
Prevents the pump from constantly
starting and stopping when the electrode
detects water.
4 0 – 10 sec. 3 sec.
(on devices
of < 65 litres
0 sec.)
77 FlowSwChckT
(Global 5)
Time after which the water feed pump or
vacuum pump switches off if no water is
detected by the flow sensor.
4 1 – 30 sec. 8 sec.
78 FlowTryTime
(Global 6)
Time period until the water feed pump or
vacuum pump will restart after detecting
a water shortage.
4 0 – 9000 sec. 30 sec.
79 VacuumPump
(Global 10)
This parameter is set to 3 in all devices
and should not be viewed in connection
with a vacuum unit.
Exception:
The value is set to 10: The vacuum unit
pumps out the accrued condensation
during the sterilisation phase at
temperatures under 100 °C SterTemp.
This prevents the sterilisation chamber
from filling up completely.
4 0 – 9999 3
80 HeatTime1
(Cycle 0)
Ramp function:
Heating time until the first ramp (or the
temperature selected in phase 1) is
reached.
Ramp function
0
50
100
150
Temperature °C
Phase 1 Phase 2 Phase 3
4 0 – 9999 min. 0 min.
81 StayTime1
(Cycle 1)
Ramp function:
Time for which the temperature set for
phase 1 is held.
4 0 – 9999 sec. 0 sec.
82 StayTemp1
(Cycle 2)
Ramp function:
“Hold temperature” for the 1st ramp.
4 3.0 – 130.0 °C 3.0 °C
83 StayPrsAdd1
(Cycle 3)
Ramp function:
Additional pressure to the corresponding
steam pressure according to
“StayTemp1”
4 0.0 – 200.0 kPa 40.0 kPa
84 HeatTime2
(Cycle 4)
Ramp function:
Heating time until the 2nd ramp (or the
temperature selected in phase 2) is
reached.
4 0 – 9999 min. 0 min.
85 StayTime2
(Cycle 5)
Ramp function:
Time for which the temperature set for
phase 2 is held.
4 0 – 9999 sec. 0 sec.