V/D
S
ERIES
P
ARAMETERS
Revision: 2.1 8-11
No.
Parameter Description Access
level
Range of values Factory
setting
86 StayTemp2
(Cycle 6)
Ramp function:
“Hold temperature” for the 2nd ramp.
4 3.0 – 130.0 °C 3.0 °C
87 StayPrsAdd2
(Cycle 7)
Ramp function:
Additional pressure to the corresponding
steam pressure according to
“StayTemp2”.
4 0.0 – 200.0 kPa 40.0 kPa
88 HeatSterTime
(Cycle 8)
Ramp function:
Heating time until the sterilisation
temperature is reached (phase 3 =
sterilisation phase).
4 0 – 3600 min. 0 min.
89 RemoveWaterT
(Cycle 41)
For spray cooling or
Eco/Basic devices, for example:
Run-on time at the end of the program
for the drain valve after the water level
electrode can no longer detect water in
the chamber.
4 0 – 9999 sec. 30 sec.
90 VacDelay
(Cycle 43)
Run-on time before the vacuum valve
opens after the vacuum pump starts.
4 0 – 60 sec. 1 sec.
91 VacWaitTime
(Cycle 44)
For solids cycles with quick steam
release:
Run-on time before the vacuum pump
starts after reaching 115 kPa.
4 0 – 9999 sec. 5 sec.
92 Hot Water
(Cycle 69)
Hot water:
0 = Steam sterilisation
1 = Steam air mixture or hot water
sprinkling
4 0, 1 0
93 Water Time
(Cycle 72)
X devices:
0 = No water in the sterilisation chamber
> 0 = Run-on time where the water inlet
valve stays open after the water level
electrode has detected water in the
sterilisation chamber (e.g. spray
cooling)
Eco/Basic devices:
0 = Automatic filling – also during the
sterilisation phase (long-term
sterilisation, material tester)
1 = Manual filling with water disposal
after the end of the program
(demineralised water is drained out
of the sterilisation chamber)
2 = Manual filling without water disposal
after the end of the program when
the EndTemp ≤ 80 °C.
(demineralised water is retained in
the sterilisation chamber for the next
program cycle)
> 2 = Run-on time where the water inlet
valve stays open after the water level
electrode (E3) has detected water in
the sterilisation chamber
4 0 – 300 sec. 0 sec.