Page 6 of 6
13 Tighten the lockdown cap into the top of the furnace lift shaft. This immobilizes the furnace.
Figure 12
14 On the Q400 display panel, press the Control button.
15 Press the Shutdown button, then press Apply. Wait for the display screen to show that the instrument
is shutdown and it is safe to turn off the instrument.
16 Turn off the power switch located on the back of the instrument near the power cord.
17 Bolt the Q400 down on the shipping pallet. Place the cardboard box over the pallet.
18 Strap the cardboard box down over the pallet and instrument.
19 Seal the box and place the green return label on the top of the box.