Issued June 2017 Page 1 of 6
TA Instruments
Packing the TMA Q400
This procedure covers the steps for properly packing up the TMA Q400 for shipment.
Packing Procedure
1 Remove the probe and stage from the instrument
2 Remove the outer cover around the shuttle assembly.
3 Remove the weight tray. Tape the nylon screw in place and lay the weight tray in the top of the furnace.
See Figures 1 and 2.
Figure 1