5. Begin by securing the bridge piece to the bath. Use the two M4 countersunk stainless screws and screw into
the holes in the corners of the bath.
6. Locate the thermoregulator above the bridge piece. Toggle the heating element carefully through the
clearance hole. Lower the thermoregulator onto the M4 threaded stud (1) and 3mm diameter stud (2). Use the
half-turn plastic fastener (3) to clamp down the thermoregulator.
Use M4 countersunk
screws
1
3
5
6
4
2