CSBCryoEMFacilityTF20OperatingInstructionsSEC
05/02/16
VII. EndofSession
14. Closecolumnvalves
15. Retractthecampera
16. Putthescreendown
17. Home the stage (go to the Stage tab Control and click on Holder, this ensures that the stage
isat0uminX,YandZ.
18. Removetheholdercarefully.
19. Clickoncryocycle.Makesurethecryocyclewillrunfor365minutesormore.
Wait~5minutesorforTurbototurnyellowbeforeremovingtheliquidnitrogendewar
20. Remove the dewar from the microscope and dispose of the remaining LN2 in the foam
containerunderthetable.
21. Placetheprotectivefabricunderneaththeanticontaminator.
22. Dismountthespecimenfromtheholderanddisposeofit(orsaveit)properly.
23. IfusingtheRTorRTtomographyholders,returnthemtotheirproperreceptacle.
24. If using the Gatan cryo holder(s), return to the dry pumping station and follow the instructions
thereonhowtowarmitandpumpitproperly.
25. CloseDigitalMicrographandtheTecnaiUserInterfaceandlogout.
26. Pleasesignupontheuser'ssheet.Recordtheactualtimespentimaging.
27. Return the next day to retrieve Data. Data will be moved overnight to the computer located
behindthecurtainintheF20room