EasyManua.ls Logo

Tektronix 2213 - Page 63

Tektronix 2213
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance
C
heck
Procedure-2213
Service
k
.
R
epeat
p
art
d
.
2
.
Check
External
Triggering
α
.
Set
:
4-
1
0
VOLTS/DIV
(both)
10
mV
VERTICAL
MODE
C
H
1
SEC/DIV
20
μ
s
b
.
Connect
the
test
setu
p
as
shown
in
Figure
4-1
.
F
igure
41
.
Test
setup
for
external
trigger
and
jitter
c
hecks
.
e
.
M
ove
the
signal
from
the
CH
1
OR
Χ
input
connector
to
the
ΕΧΤ
INPUT
connector
.
f
.
Set
the
generator
to
2
MHz
.
g
.
CHECK-Stable
dis
p
lay
can
be
obtained
by
adjusting
the
TRIGGER
LEVEL
control
for
each
switch
combination
given
in
Table
4-6
.
h
.
R
emove
the
10
Χ
attenuator
from
the
test
setup
and
set
the
ΕΧΤ
COUPLING
switch
to
DC
,
10
.
LEVELED
SI
NE-W
A
VE
G
ENER
ATO
R
3827-10
REV
NOV
1981
h
.
Set
the
generator
output
to
produce
α 1
.5-
d
ivision,
60-MHz
d
is
p
lay
.
c
.
Set
the
leveled
5-division,
50-kHz
display
.
sine-wave
generator
to
p
roduce
α
i
.
R
ep
eat
p
art
d
.
d
.
Set
:
VERTICAL
MODE
CH
2
j .
Move
the generator
output
from
the
CH
2
OR
Υ
SEC/DIV
0
.2
μ
S
input
connector
to
the
CH
1
OR
Χ
input
connector
.
Set
T
RIGGER
SOURCE
ΕΧΤ
VERTICAL
M
ODE
to
CH
1
.

Other manuals for Tektronix 2213