EasyManua.ls Logo

Tektronix 2221 - APPENDIX A: PERFORMANCE CHECK PROCEDURE; TEST EQUIPMENT REQUIRED, LIMITS AND TOLERANCES, PREPARATION FOR CHECKS

Tektronix 2221
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
A
-
2221
Operators
APPENDIX
A
PERFORMANCE
CHECK
PROCEDURE
П
Л
П
П
П
П
П
INTRODUCTION
PURPOSE
The
Performance
Check
Procedure
is
used
to
verify
"
Performance
Requirements
"
listed
in
Table
1-2
and
to
determine
the
need
for
readjustment
.
The
performance
checks
may
also
be
used
as
an
acceptance
test
or
as
a
preliminary
troubleshooting
aid
.
the
minimum
necessary
to
provide
accurate
results
.
There
fore
,
equipment
used
must
meet
or
exceed
the
listed
specifications
.
Detailed
operating
instructions
for
test
equipment
are
not
given
in
this
procedure
.
If
more
operat
ing
information
is
required
,
refer
to
the
appropriate
test
equipment
instruction
manual
.
PERFORMANCE
CHECK
INTERVAL
When
equipment
other
than
that
recommended
is
used
,
control
settings
of
the
test
setup
may
need
to
be
changed
.
If
the
exact
item
of
equipment
given
as
an
example
in
Table
A
-
1
is
not
available
,
check
the
"
Minimum
Specification
"
column
to
determine
if
any
other
available
test
equipment
can
be
used
to
perform
the
check
.
}
To
ensure
instrument
accuracy
,
check
its
performance
after
every
2000
hours
of
operation
(
or
once
each
year
if
used
infrequently
)
.
A
more
frequent
interval
may
be
neces
sary
if
the
instrument
is
subjected
to
harsh
environments
or
severe
usage
.
LIMITS
AND
TOLERANCES
STRUCTURE
This
Performance
Check
Procedure
is
structured
in
subsections
to
permit
checking
individual
sections
of
the
instrument
whenever
a
complete
Performance
Check
is
not
required
.
At
the
beginning
of
each
subsection
there
is
an
equipment
-
required
list
showing
only
the
test
equip
ment
necessary
for
performing
the
steps
in
that
subsec
tion
.
In
this
list
,
the
Item
number
that
follows
each
piece
of
equipment
corresponds
to
the
Item
number
listed
in
Table
A
-
1
.
The
tolerances
given
in
this
procedure
are
valid
for
an
instrument
that
is
operating
in
and
has
been
previously
calibrated
in
an
ambient
temperature
between
+
20
°
C
and
+
30
°
C
.
The
instrument
also
must
have
had
at
least
a
20
minute
warm
-
up
period
.
Refer
to
Table
1-2
for
tolerances
applicable
to
an
instrument
that
is
operating
outside
this
temperature
range
.
All
tolerances
specified
are
for
the
instrument
only
and
do
not
include
test
-
equipment
error
.
PREPARATION
FOR
CHECKS
n
n
П
n
П
Л
7
n
Also
at
the
beginning
of
each
subsection
is
a
list
of
all
the
front
-
panel
control
settings
required
to
prepare
the
instrument
for
performing
Step
1
in
that
subsection
.
Each
succeeding
step
within
a
particular
subsection
should
then
be
performed
,
both
in
the
sequence
presented
and
in
its
entirety
,
to
ensure
that
control
-
setting
changes
will
be
correct
for
the
steps
that
follow
.
It
is
not
necessary
to
remove
the
instrument
cover
to
complete
any
of
the
subsections
in
the
Performance
Check
Procedure
,
since
all
checks
are
made
using
operator
-
accessible
front-
and
rear
-
panel
controls
and
connectors
.
TEST
EQUIPMENT
REQUIRED
The
most
accurate
display
adjustments
are
made
with
a
stable
,
well
-
focused
,
low
-
intensity
display
.
Unless
other
wise
noted
,
adjust
the
INTENSITY
,
STORE
/
READOUT
INTENSITY
,
FOCUS
,
and
TRIGGER
LEVEL
controls
as
needed
to
view
the
display
.
Table
A
-
1
is
a
complete
list
of
the
test
equipment
required
to
accomplish
the
Performance
Check
Procedure
.
Test
equipment
specifications
described
in
Table
A
-
1
are
o
A
-
1

Related product manuals