Basic
Applications
2221
Operators
MAKING
NONSTORAGE
MEASUREMENTS
2
The
following
procedures
will
enable
the
operator
to
perform
some
basic
measurements
and
familiarize
the
operator
with
the
conventional
oscilloscope
capabilities
of
the
2221
.
AC
PEAK
-
TO
-
PEAK
VOLTAGE
To
make
a
peak
-
to
-
peak
voltage
measurement
,
use
the
following
procedure
:
POSITION
TO
CENTERLINE
B
100
NOTE
90
This
procedure
may
also
be
used
to
make
voltage
measurements
between
any
two
points
on
the
waveform
.
U
A
VERTICAL
DEFLECTION
10
!
07
.
V
u
1.
Preset
instrument
controls
and
obtain
a
baseline
trace
.
5V
50us
MEASURE
AMPLITUDE
FROM
(
A
)
TOB
2.
Apply
the
ac
signal
to
either
vertical
-
channel
input
connector
and
set
the
VERTICAL
MODE
Switches
to
display
the
channel
used
.
4998-30
بألا
Figure
6-1
.
Peak
-
to
-
peak
waveform
voltage
.
3.
Set
the
appropriate
VOLTS
/
DIV
switch
to
display
about
five
divisions
of
the
waveform
,
ensuring
that
the
VOLTS
/
DIV
Variable
control
is
in
the
CAL
detent
.
NOTE
4.
Adjust
the
TRIGGER
LEVEL
control
to
obtain
a
stable
display
.
A
more
accurate
value
can
be
obtained
by
measur
ing
from
the
top
of
a
peak
to
the
top
of
a
valley
.
This
eliminates
trace
thickness
from
the
measure
ment
.
u
5.
Adjust
the
SEC
/
DIV
switch
to
display
several
cycles
of
the
waveform
.
9.
Calculate
the
peak
-
to
-
peak
voltage
,
using
the
follow
ing
formula
:
6.
Vertically
position
the
displayed
waveform
so
that
the
negative
peak
of
the
waveform
coincides
with
one
of
the
horizontal
graticule
lines
(
see
Figure
6-1
,
Point
A
)
.
Volts
(
p
-
p
)
Vertical
VOLTS
/
DIV
Probe
Deflection
X
Switch
X
Attenuation
(
Divisions
)
Setting
Factor
7.
Horizontally
position
the
display
so
that
one
of
the
positive
peaks
coincides
with
the
center
vertical
graticule
line
(
see
Figure
6-1
,
Point
B
)
.
EXAMPLE
:
The
measured
peak
-
to
-
peak
vertical
deflection
is
5
divisions
(
see
Figure
6-1
)
with
a
VOLTS
/
DIV
switch
setting
of
0.5
V
,
using
a
10X
probe
.
Substituting
the
given
values
:
8.
Measure
the
vertical
deflection
from
peak
to
peak
(
see
Figure
6-1
,
Point
A
to
Point
B
)
.
Volts
(
p
-
p
)
=
5
div
x
0.5
V
/
div
X
10
=
25
V
6-4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2221 and is the answer not in the manual?

Tektronix 2221 Specifications

General IconGeneral
BrandTektronix
Model2221
CategoryTest Equipment
LanguageEnglish

Related product manuals