EasyManuals Logo

Tektronix 2445 User Manual

Tektronix 2445
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
Operator's Checks and Adjustments—2445 Operators
3. Connect the Channel 1 probe (using the probe hook
tip) to the oscilloscope CALIBRATOR output.
4. Set triggering controls for a stable display. The dis
play should be five cycles of the CALIBRATOR square-
wave signal, with an amplitude of four divisions. Center the
display on the screen.
5. Check the waveform for overshoot and rolloff (see
Figure 5-1). If necessary, use a small-bladed screwdriver to
adjust the probe low-frequency compensation for a square
front corner on the waveform.
6. Release the CH 1 VERTICAL MODE switch, select
CH 2 VERTICAL MODE, and connect the Channel 2 probe
input to the CALIBRATOR output. Observe that the
CH 2 SCALE FACTOR on the readout display indicates
100 mV with the 10X probe attached.
7. Use the Channel 2 POSITION control to vertically
center the display and repeat step 5 for the Channel 2
probe.
NOTE
Refer to the instruction manual supplied with the
probe for more complete information about low-
frequency and high-frequency compensation o f
the probes.
MATCHING CHANNEL 2 DELAY
The apparent signal delay in Channel 2 may be adjusted
up to ±500 ps to match the apparent delay present in any
of the other three channels. This adjustment is most com
monly used to eliminate delay differences between Chan
nel 1 and Channel 2 that may be introduced by the probes
and has no effect on common-mode rejection when ADD
VERTICAL MODE is selected. Matching Channel 1 and
Channel 2 delay is accomplished as follows:
1. Preset the instrument controls and obtain a baseline
trace as described in "Initial Setup."
2. Connect the two 10X probes (supplied with the
instrument) to the CH 1 OR X and CH 2 input connectors.
3. Check and adjust, if necessary, the probes' low-
frequency compensation. Refer to "Probe Low-Frequency
Compensation" in this section.
4. Connect both probes via hook tips to a pulse gener
ator fast-rise output.
5. Press in both the CH 1 and CH 2 VERTICAL MODE
switches.
6. Set oscilloscope triggering controls for a stable
display.
Figure 5-1. Probe low-frequency compensation.
7. Set the A AND B SEC/DIV switches to 10 ns.
8. Adjust the Channel 1 and Channel 2 POSITION
controls to vertically overlay the two displayed signals.
9. Pull out the B SEC/DIV switch and observe the
message CH 2 DELAY TURN A in the upper right-hand
corner of the screen.
NOTE
The 2445 can be set to preclude operator adjustment
of Channel 2 delay. If the delay-offset feature is
disabled, the message CH 2 DLY DISABLED appears
in the top row of the readout when attempting to
activate the feature. Refer the instrument to a
qualified service technician if adjustment of the
delay matching is required.
5-3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2445 and is the answer not in the manual?

Tektronix 2445 Specifications

General IconGeneral
BrandTektronix
Model2445
CategoryTest Equipment
LanguageEnglish

Related product manuals