How to sanitize a nonfunctional instrument
The end of the memory card lifts upward as you remove the screw.
7.
Grasp the edges of the raised end of the card and pull to remove the memory card. Store or destroy the memory
card in accordance with your organization’s guidelines.
8.
Reattach the d rive cover.
9.
Package the instrument for shipping. Contact Tektronix for guidelines on correct packaging to best protect your
instrument during shipping.
10.
Send the instrument to a Tektronix Service Center. The instrument will then be repaired, calibrated as necessary,
andreturnedtoyou.
14 MSO58LP with Option 5-SEC Declassication and Security Instructions