f. Check against limits:
CHECK that the readout for the measurement Mean readout on
screen is within the limits listed for the current vertical scale and
termination/offset/generator settings. Enter the value on test record.
Repeat substep 2.d, reversing the polarity of the termination voltage,
offset, and generator settings as is listed in the table.
CHECK that the Mean measurement readout on screen is within the
limits listed for the current vertical scale setting and termination/
offset/generator settings. Enter the value on test record.
Repeat substeps 2.c through 2.f until all vertical scale settings are
checked for the channel under test. (See Table 6: DC voltage
measurement accuracy on page 178.)
g. Test all channels: Repeat substeps 2.a through 2.f for all four channels.
3. Disconnect the hookup:
a. Set the generator output to 0 V.
b. Disconnect the equipment from the generator output and the input
connector of the channel last tested.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
182 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals