CHECK that the difference obtained is within limits as follows:
Model Limits
≥ 4 GHz models 343.2 mV to 536.8 mV
< 4 GHz models 985 mV to 1015 mV
Enter voltage difference on test record.
3. Disconnect the hookup: Disconnect the equipment from Ch 1.
Serial trigger checks (Optional on some models)
These procedures check those characteristics that relate to the serial trigger
system and are listed as checked in Specifications.
Check serial trigger baud rate limits
Equipment required Prerequisites
One precision 50 Ω coaxial cable (Item 4)
One sine-wave generator (Item 9)
One adapter (Item 19)
(See Prerequisites on page 116.) Also, the
instrument must have passed Check DC Voltage
Measurement Accuracy. (See Signal acquisition
system checks on page 176.)
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)