EasyManuals Logo
Home>Tektronix>Test Equipment>DPO5000 Series

Tektronix DPO5000 Series Technical Reference

Tektronix DPO5000 Series
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #335 background imageLoading...
Page #335 background image
Gain Accuracy = ((Vdiff- VdiffExpected)/VdiffExpected) × 100%
d. Enter Gain Accuracy in the worksheet and in the test record.
10. Repeat tests at 1 MΩ impedance as follows:
a. Set the calibrator to 1 MΩ output.
b. Push the front-panel Vertical V menu button.
c. Set the Termination to 1 MΩ
d. Repeat steps 8 through 9.
11. Repeat steps 7 through 10 for all vertical scale settings in the test record.
12. Repeat the procedure for all remaining channels:
a. Push the front-panel button to deselect the channel that you have already
tested.
b. Push the front-panel button to select the next channel to be tested.
c. Move the DC voltage source connection to the channel input to be tested.
d. Starting from step 3, repeat the procedure until all channels have been
tested.
13. Press the Menu Off button.
Check offset accuracy
This test checks the offset accuracy.
Performance verification (MSO/DPO5000/B series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 323

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals