Vertical scale Reference
amplitude
Horizontal scale Test frequency -3 dB
DPO7254C DPO7104C DPO7054C
10 mV 60 mV 1 ns 2.501 GHz 1 GHz 500 MHz ≥42.4 mV
20 mV 120 mV 1 ns 2.501 GHz 1 GHz 500 MHz ≥84.8 mV
50 mV 300 mV 1 ns 2.501 GHz 1 GHz 500 MHz ≥212 mV
100 mV 600 mV 1 ns 2.501 GHz 1 GHz 500 MHz ≥424 mV
200 mV 1.2 V 1 ns 2.501 GHz 1 GHz 500 MHz ≥848 mV
500 mV 3 V
7
1 ns 2.501 GHz 1 GHz 500 MHz ≥2.12 V
7
1 V 5 V
7
1 ns 2.501 GHz 1 GHz 500 MHz ≥3.535 V
7
c. Set the vertical scale:
For the channel you are testing, set the vertical Scale to the next setting
in the table, starting with the 100 mV setting. (See Table 12: Analog
bandwidth, < 3.5 GHz models on page 228.)
d. Set thetriggering coupling: From the button bar, touch Trigger, then
select Coupling DC.
e. Display the test signal: Do the following subparts to first display the
reference signal and then the test signal.
From the button bar touch Measure; then select the Time tab.
Touch the Freq button to measure the frequency of the current
channel.
Select the Ampl tab. Touch the Amplitude button.
Touch the Statistics button, then touch the Mean button.
Touch the X (Close) button.
Set the generator output so the Chx Amplitude mean readout equals
the reference amplitude in the table. This corresponds to the vertical
scale set in substep 2.c. (See Table 12: Analog bandwidth, <
3.5 GHz models on page 228.)
Record the reading on the power meter.
Set the trigger as follows:
Press the front-panel Push-Set 50% as necessary to trigger a stable
display. At full bandwidth, you may also want to make small, manual
adjustments to the trigger level. You can use the Trigger Level knob to
do this.
f. Measure the test signal:
Set the frequency of the generator, as shown on screen to the test
frequency in the table that corresponds to the vertical scale set in substep
7
If your generator cannot output the required amplitude, determine its maximum output at the Test frequency, and use this for the reference amplitude.
The -3 dB limit can be calculated as: 0.707 × reference amplitude.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 229

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals