6. Disconnect the hookup:
a. Disconnect all test equipment from the instrument.
b. Set the instrument controls:
From menu mode, touch Utilities and select External Signals.
Touch the Internal button to select the internal reference.
c. Perform a signal path compensation:
Touch Utilities and select Instrument Calibration.
Touch Run SPC (Calibrate on some instruments) and wait for the
signal path compensation to finish.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 245