Set the Vertical Scale to 500 mV /div (≤20 GHz models) or
120 mV/div (>20 GHz models).
Use the Vertical Position knob to center the display on screen.
b. Measure logic levels:
From the button bar, touch Measure and select the Ampl tab.
Touch the High and Low buttons.
Touch the X (close) button.
c. Check Aux Out output against limits: (See Figure 36: Measurement of
trigger out limits on page 267.)
≤20 GHz models: CHECK that the Ch 1 High readout is ≥1.0 volt
and that the Ch 1 Low readout ≤0.25 volts.
>20 GHz models: CHECK that the Ch 1 High readout is ≥0.5 volt
and that the Ch 1 Low readout ≤0.125 volts.
3. Disconnect the hookup: Disconnect the test setup from the inputs and
outputs.
Figure 36: Measurement of trigger out limits