3. Record the reference level: Note the reading on the level meter.
4. Set the generator to the new frequency and reference level:
■
Change the sine wave generator to the desired new frequency.
■
Input the correction factor and/or the new frequency into the level meter.
■
Adjust the sine wave generator amplitude until the level meter again
reads the value noted in step 3. The signal amplitude is now correctly set
for the new frequency.