Random Noise, Sample Acquisition Mode
Performance checks Vertical sensitivity = 100 mV/div
1 MΩ 50 Ω
Bandwidth Test result (mV) High limit (mV) Test result (mV) High limit (mV)
Channel 3 Full 8.13 8.13
250 MHz limit 6.10 6.10
20 MHz limit 4.10 4.10
Channel 4 Full 8.13 8.13
250 MHz limit 6.10 6.10
20 MHz limit 4.10 4.10
Random Noise, Sample Acquisition Mode
Performance checks Vertical sensitivity = 100 mV/div
1 MΩ 50 Ω
Bandwidth Test result (mV) High limit (mV) Test result (mV) High limit (mV)
MSO5204B, DPO5204B
Channel 1 Full 5.0 4.5
250 MHz limit 3.8 3.7
20 MHz limit 4.0 3.0
Channel 2 Full 5.0 4.5
250 MHz limit 3.8 3.7
20 MHz limit 4.0 3.0
Channel 3 Full 5.0 4.5
250 MHz limit 3.8 3.7
20 MHz limit 4.0 3.0
Channel 4 Full 5.0 4.5
250 MHz limit 3.8 3.7
20 MHz limit 4.0 3.0
MSO5104B, DPO5104B
Channel 1 Full 4.5 4.5
250 MHz limit 3.8 3.7
20 MHz limit 4.0 3.0
Channel 2 Full 4.5 4.5
250 MHz limit 3.8 3.7
20 MHz limit 4.0 3.0
Channel 3 Full 4.5 4.5
250 MHz limit 3.8 3.7
20 MHz limit 4.0 3.0