Theory of Operation
Power System
A 250 W power sup
ply generates a switched 12 volt bulk supply as well as a
5 volt standby supply.
No switch comp
letely disconnects t he line power from the instrument. The power
switch controls the power to the instrument through the COM-E board circuitry.
When the instrument is in the "power off" condition, there is still a low power
standby circuit to allow the system to monitor the power switch.
Fans
There are two instrument fans: the sys tem fan, which is mounted on the rear
chassis, and a smaller fan, which is mounted on the COM-E module. For cooling,
the system fan draws air into the instrumen t, for a positively pressured instrument.
ThesystemfaniscontrolledbyanembeddedprocessorontheMainboardandis
regulated by monitoring the temperature of the Analog acquisition board circuitry.
Cooling
air enters the instrument through speci c locations on the covers, where
it ows directly over the heat s inks of each high wattage component. The COM-E
fan cools the COM-E m odule.
2–12 MSO70000/C, DSA70000B/C, DPO70000B/C, DPO7000, MSO5000, DPO5000 Series

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals