Environmental specifications
Temperature <4 GHz models
Operating +5 °C to +45 °C (41 °F to +113 °F), with noncondensing conditions
Nonoperating -40 °C to +71 °C (-40 °F to +160 °F), with 15 °C/hour maximum gradient, without disk media
installed in disk drives
Temperature ≥4 GHz models
Operating +5 °C to +45 °C (41 °F to +113 °F), with 11 °C per hour maximum gradient, noncondensing,
derated 1 °C per 300 meters (984.25 feet) above 1500 meters (4921.25 feet) altitude
Nonoperating -20 °C to +60 °C (-4 °F to +140 °F), with 20 °C/hour maximum gradient, without disk media
installed in disk drives
Humidity >4 GHz models
Operating 8% to 80% relative humidity with a maximum wet-bulb temperature of +29 ºC (84 °F) at or below
+45 ºC (113 °F), noncondensing
Upper limit derated to 30% relative humidity at +45 ºC (+113 °F)
Nonoperating 5% to 90% relative humidity with a maximum wet-bulb temperature of +29 ºC (84 °F) at or below
+60 ºC (140 ºF), noncondensing
Upper limit derated to 20% relative humidity at +60 ºC (+140 °F)
Humidity ≥4 GHz models
Operating 8% to 80% relative humidity at up to +32 °C (+90 °F)
5% to 45% relative humidity above +32 °C (+90 °F) up to +45 °C (+113 °F), noncondensing, and is
limited by a maximum wet-bulb temperature of +29.4 °C (+85 °F) (derates relative humidity to 32%
at +45 °C (+113 °F)
Nonoperating 5% to 95% relative humidity at up to +30 °C (+86 °F),
5% to 45% relative humidity above +30 °C (+86 °F), up to +60 °C (+140 °F), noncondensing, and
is limited by a maximum wet-bulb temperature of +29.4 °C (+85 °F) (derates relative humidity to
11% at +60 °C (+140 °F))
Altitude <4 GHz models
Operating Up to 3,000 meters (9,843 feet)
Nonoperating Up to 12,192 meters (40,000 feet)
Altitude ≥4 GHz models
Operating Up to 3,000 meters (9,843 feet), derate maximum operating temperature by 1 °C per 300 meters
(984.25 feet) above 1500 meters (4921.25 feet) altitude
Nonoperating Up to 12,000 meters (39,370 feet)
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series)
76 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series

Table of Contents

Other manuals for Tektronix DSA70000D Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000D Series and is the answer not in the manual?

Tektronix DSA70000D Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000D Series
CategoryTest Equipment
LanguageEnglish

Related product manuals