FAQs
How should I clean the front‑panel display?
If you need to clean the front-panel LCD touchscreen display,
use a soft dry cloth. If necessary, moisten a microber cloth
with distilled water or diluted isopropyl alcohol and gently
wipe the display. Do not spray or apply uids directly to the
LCD display. Additionally, it is strongly recommended that
you use only ngers on the touchscreen. Do not use sharp
objects, such as a pen or pencil, to touch the touchscreen.
My data look odd or is wrong. What should I do?
Verify the connections from the instrument to the test xture.
Also, check the connections from the DUT to the test xture
socket.
I cannot unplug the mini‑triaxial cable (4200A‑MTRX)
from the SMU. What should I do?
The mini-triaxial connector is a locking connector. To remove
it, pull the knurled part of the connector back.
A conguration error has been detected and I cannot
launch Clarius. What should I do?
This occurs when the physical conguration does not
match the conguration dened in KCon or when there are
communication problems between instruments and the
4200A-SCS.
To verify system conguration:
1. Run KCon.
2. Select Validate.
To update the system conguration:
1. Run KCon.
2. Select Update.
This can occur if the preamplier or RPM is removed or
reconnected. Note that preampliers are SMU specic. For
example, a preamplier that is congured for SMU1 cannot
be connected to SMU2.