FAQs
How should I clean the front‑panel display?
If you need to clean the front-panel LCD touchscreen display,
use a soft dry cloth. If necessary, moisten a microber cloth
with distilled water or diluted isopropyl alcohol and gently
wipe the display. Do not spray or apply uids directly to the
LCD display. Additionally, it is strongly recommended that
you use only ngers on the touchscreen. Do not use sharp
objects, such as a pen or pencil, to touch the touchscreen.
My data look odd or is wrong. What should I do?
Verify the connections from the instrument to the test xture.
Also, check the connections from the DUT to the test xture
socket.
I cannot unplug the mini‑triaxial cable (4200A‑MTRX)
from the SMU. What should I do?
The mini-triaxial connector is a locking connector. To remove
it, pull the knurled part of the connector back.
A conguration error has been detected and I cannot
launch Clarius. What should I do?
This occurs when the physical conguration does not
match the conguration dened in KCon or when there are
communication problems between instruments and the
4200A-SCS.
To verify system conguration:
1. Run KCon.
2. Select Validate.
To update the system conguration:
1. Run KCon.
2. Select Update.
This can occur if the preamplier or RPM is removed or
reconnected. Note that preampliers are SMU specic. For
example, a preamplier that is congured for SMU1 cannot
be connected to SMU2.