Acquire the Signal
Using MagniVu
1. Push D15 D0.
2. Push MagniVu and select On.
D15–D0
On/Off
Thresholds Label MagniVu
On |
Off
Height
S |ML
Quick Tips
If you think you need more timing resolution, turn on MagniVu to increase the resolution.
MagniVu is always acquired. If the oscilloscope is in a stoppe d s tate, you can turn on MagniVu and still get th e resolution
without taking another acquisition.
The serial bus features do not use data acquired in MagniVu mode.
MSO4000 and DPO4000 Series Oscilloscopes User Manual 129