Check that oscilloscope passes power-on self tests
Power-on self tests verify that all oscilloscope modules are working correctly after power up.
1. Power on the oscilloscope and wait until the oscilloscope screen appears.
2. Select Utility > Self Test from the top-edge Menu bar to open the Self Test configuration menu.
3. Check that the status of all power-on self tests are Passed.
If one or more power-on self tests shows Failed:
a. Power cycle the oscilloscope.
b. Tap Utility > Self Test. If one or more power-on self tests still shows Failed, contact Tektronix Customer Support.
Connecting Probes
Probes and cables connect the oscilloscope to your device under test (DUT). Use a probe that best matches your signal
measurement needs.
Figure 2: Connecting probes to the MSO44 or MSO46
Connect TPP Series, TekVPI+, TekVPI, or other supported Tektronix analog probes by pushing them into a FlexChannel
connector. The probe base latch locks with a 'click' when the probe is fully seated.
TekVPI probes automatically set the channel input parameters for that probe (bandwidth, attenuation, termination, and so on). If
a probe has a Menu button, push that button to open an on-screen configuration menu. Follow instructions provided with active
probes to set their parameters (auto zero, degauss, and so on).
To connect a TLP058 FlexChannel Logic Probe or a TDP7700 Series TriMode
Probe:
1. Move the locking lever to the unlocked position, then let go to reset locking lever to the center position.
2. Insert the probe into a FlexChannel connector until fully seated and the lock mechanism clicks.
Installing your instrument
MSO44, MSO46 Installation and Safety Manual 5
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com

Table of Contents

Other manuals for Tektronix MSO44

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO44 and is the answer not in the manual?

Tektronix MSO44 Specifications

General IconGeneral
Bandwidth200 MHz, 350 MHz, 500 MHz, 1 GHz
Analog Channels4
Operating Temperature0 °C to +50 °C
Sample Rate6.25 GS/s
Record Length62.5 Mpoints
Input Impedance1 MΩ ±1%, 50 Ω ±1%
Vertical Resolution12 bits
Waveform Capture RateUp to 500, 000 wfm/s
Maximum Input Voltage5 VRMS (50 Ω)
Probe InterfaceTekVPI
Digital Channels16 (optional)