h. Repeat substeps 2.d through 2.g, using the next position, offset and
generator settings listed in the table for the current vertical scale.
i. Repeat substeps 2.c through 2.h until all vertical scale settings, listed in
the table, are checked for the channel under test. (See Table 8: Gain
accuracy on page 189.)
j. Test all channels: Repeat substeps 2.a through 2.i for all four channels.
3. Disconnect the hookup:
a. Set the generator output to 0 V.
b. Disconnect the generator output from the channel last tested.