Specifications
Duty cycle 0.0
007
t
meas
–t
reference
10 ms
Measurement time position excludes the beginning and ending of the pulse extending for a time = (10/measurement Bandwidth) as measured over the
center 50% of the pulse.
4
For conditions of:
CW (non-chirped) pulses
Frequency Estimation = Manual
Pulse ON power –20 dBm
Signal peak at Ref Lvl.
Atten=0dB
Pulse width 2 s
PRI 300 us.
Duty cycle 0.001
t
meas
–t
reference
10 ms
Measurement time position excludes the beginning and ending of the pulse extending for a time = (10/measurement Bandwidth) as measured over the
center 50% of the pulse.
5
For conditions of:
Linear Chirped pulses
For signal type: Linear Chirp, Peak-to-peak Chirp Deviation: 0.8 x Measurement BW.
Frequency Estimation = Manual
Pulse ON power –20 dBm
Signal peak at Ref Lvl.
Atten=0dB
Pulse width 100 ns.
PRI 300 us.
Duty cycle 0.0003
t
meas
–t
reference
10 ms
Measurement time position excludes the beginning and ending of the pulse extending for a time = (10/measurement Bandwidth) as measured from the
50% point of the Tr or Tf.
6
For conditions of:
Linear Chirped pulses
For signal type: Linear Chirp, Peak-to-peak Chirp Deviation: 0.8 x Measurement BW.
Frequency Estimation = Manual
Pulse ON power –20 dBm
Signal peak at Ref Lvl.
Atten=0dB
Pulse width 300 ns.
PRI 1000 us.
Duty cycle 0.0003
t
meas
–t
reference
10 ms
Measurement time position excludes the beginning and ending of the pulse extending for a time = (10/measurement Bandwidth) as measured from the
50% poi
ntoftheTrorTf.
7
For conditions of:
CW (non-chirped) pulses
Frequency Estimation = Manual
Pulse ON power –20 dBm
Signal peak at Ref Lvl.
Atten=0dB
Pulse width 200 ns.
PRI 300 us.
Duty cycle 0.0007
t
meas
–t
reference
10 ms
Measurement time position excludes the beginning and ending of the pulse extending for a time = (10/measurement Bandwidth) as measured from the
50% point of the Tr or Tf.
22 RSA5100B Series Technical Reference