Service
Information-T921/T922/T922R
2
.
Use
a
small
screwdriver
to
remove
the
clip
holding
Q458
to
the
rear
subpanel
.
5-
1
6
3
.
Remove
the
front
panel
knobs
and
shafts
.
4
.
Remove
the
high-voltage
shield
and
the
two
posts
underneath
the
shield
.
5
.
Disconnect
J138
(white-blue
wire),
J148
(white-
brown
wire),
J470
(a
brown
two-lead
connector
and
a
black
two-lead
connector),
J475, J465, J466,
and
the
Cal
Out
lead
(goes
to
PROBE
ADJ
on
front
panel)
.
The
J138
and
J148
wires
pass
through
the
board
to
the
crt
.
NOTE
To
disconnect
J475, J466,
and
J465,
lift
the
cable
retainers
with
a
screwdriver
until
you
can
remove
the
cable
.
Also
remove
the
high-voltage
lead
from
U460
to
the
crt
.
WARNING
The
crt
anode
and
the output
terminal
of
high-
voltage
multiplier
U460
may
retain
a
10,000
volt
charge
after
the
instrument
is
turned
off
.
To
avoid
electrical
shock,
ground
both
the
output
terminal
of
U460
and
the
crt
high-voltage
anode
lead
to
chassis
ground
.
6
.
Remove
the
five
retaining
screws
.
7
.
Disconnect
J7
from
the
Power
Supply
board
by
lifting
the
Interface
board
out
.
8
.
Unsolder
the
power
cord
conductors
(a
blue wire
and
a
brown
wire)
from
the
back
of
the
board
.
To
replace the
Interface
board, reverse
the
above
procedure
.
Cathode
Ray
Tube
(CRT)
Replacement
Use
care
when
handling
a
crt
.
Protective
clothing
and
safety
glasses
should
be
worn
.
Avoid
striking
it
on
any
object
which might cause
it
to
crack
or
implode
.
When
storing
a
crt,
place
it
in
a
protective
carton
or
set
it
face
down
on
a
smooth
surface
in
a
protective
location
with
a
soft
mat
under
the
faceplate
to
protect
it
from
scratches
.
To
remove
the
crt
from
the
instrument,
disconnect
the
power
cord
plug
from
the
power
input
source
and
remove
the
cabinet
halves
.
Turn
the
front
of
theinstrument
toward
you,
and
perform
the
following
steps
.
1
.
Remove
the
two
screws
holding
the
high-voltage
shield
over the
Interface
board,
and
remove
the
shield
.
WARNING
The
crt
anode
and
the
output
terminal
of
high-
voltage
multiplier
U460
may
retain
a
10,000
volt
charge
after
the
instrument
is
turned
off
.
To
avoid
electrical
shock,
ground
both
the
output
terminal
of
U460
and
the
crt
high-voltage
anode
lead
to
chassis
ground
.
2
.
Grip the
insulated
portion
of
the
anode
lead
and
disconnect
it
from
the
U460
output
terminal,
and
ground
both
terminals
to
chassis
.
Pull
the
free
end
of
the
anode
lead
out
through
the
chassis holes
.
This
lead
is
part
of
the
crt
and
is
supplied
with
the
new
crt
.
3
.
Grip the
14-pin
crt
base
socket,
and
pull
it
backward
off
the
base
of
the
crt
.
4
.
Disconnect
the
four-pin
plug
from J470
on
the
Interface
board
.
This
terminates
two
black wires
and
two
red
wires
from
the
trace
rotation
(TR
ROT)
and
Y
AXIS
controls
.
5
.
Remove
the
three
screws
holding the
Vertical
Amplifier
to
the
chassis,
and
separate
the
amplifier
from
the
Interface
board
by
pulling
downward
to
disconnect
the
04
connector
.
6
.
Disconnect
the
two
connectors
from
the
pins
on
the
left
side
of
the
crt
neck
.
These
are the
vertical
deflection
plate
leads,
the
upper
wire color
is
blue
on white
and
the
lower
wire color
is
brown
on
white
.
7
.
Disconnect
the
two
connectors
from
the
pins
on
the
bottom
of
the
crt
neck
.
These
arethe
Horizontal
Deflection
Plate
leads
;
the
left
wire
color
is
red
on
white
and
the
right
wire color
is
green
on
white
.
8
.
Carefully
raise
the
plastic
crt
front
support
ring
upward
and
to
the
right
to
disengage
the
two
buttons on
the
bottom
and
the
two
buttons
on
the
left
side
from
the
instrument chassis
.
REV
.
B
JUL
1977