DCM300 and DCM320 Performance Verification
Handheld and Benchtop Instruments Basic Service
9
Table 9: Resistance Checks
Resistance
Calibrator Output
Tolerance Low Limit Test Result High Limit
0
±
9
000 009
Buzzer must sound
12
0
±11 109 131
Buzzer must sound
1.000 k ±0.029 k 971 1.029 k
3.700 k ±0.083 k 3.617 k 3.783 k
39.00 k ±0.87 k 38.13 k 39.87 k