TAS 200 Series Performance Verification
Handheld and Benchtop Instruments Basic Service
19
6. Position the O cursor 3 divisions below the display horizontal center with the
CURSOR POSITION switch.
7. Verify that the DV
1
(2)
readout at the top-left corner of the display, reads
between 5.78 and 6.22 V.
8. Set the VERTICAL MODE switch to CH2 and repeat steps 3 through 7 for
CH 2.