Functional Test
4–10
TDS 340A, TDS 360 & TDS 380 Technical Reference
12. Push the SAVE/RECALL front-panel button.
13. Push the Recall Factory Setup main-menu button.
14. Push the OK Confirm Factory Init side-menu button. This restores the
oscilloscope settings to those before you saved the settings.
15. Push the Recall Saved Setup main-menu button.
16. Push the From File side-menu button.
17. Turn the general purpose knob to select the file to which you saved the
settings (step 10). If you used a blank floppy disk, this file is .
18. Push the Recall From Selected File side-menu button. The oscilloscope
reads the current settings from the selected file and resets its settings. The
displayed signal should show a horizontal setting of 250 ms and the trigger at
50%.
19. Disconnect the probe from the channel input and the PROBE COMP
terminal. You are done running the file system functional test.