EasyManuals Logo

Tektronix TDS 360 Technical Reference

Tektronix TDS 360
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #88 background imageLoading...
Page #88 background image
Adjustment Procedures
TDS 340A, TDS 360 & TDS 380 Technical Reference
5–5
4. Press the main menu button Recall Factory Setup; then press the side menu
button OK Confirm Factory Init.
5. Press CLEAR MENU.
6. Connect the high output of a fast–rise step generator (Item 5) to the CH 1
BNC through a 50 W coaxial cable (Item 3) and a 50 W termination (Item 2).
(See Figure 5–3.)
7. Set the calibration generator to output a high amplitude, 1 kHz signal. Set
the pulse amplitude to 25%.
8. Set the oscilloscope VOLTS/DIV to 200 mV, the SEC/DIV to 10 ms, and
adjust the pulse amplitude for a five division display.
9. Press SET LEVEL TO 50%.
10. Use the vertical POSITION control to place the top of the waveform near
center screen.
11. Set the VOLTS/DIV to 100 mV.
12. Use an adjustment tool (Item 1) to adjust the CH 1 10X capacitor for the
flattest response. (See Figure 5–3.)
13. Remove the 50 W termination from the setup.
14. Set the VOLTS/DIV to 2 V.
10X adjustment 100X adjustment
Front of instrument
View from bottom of instrument
CH 2
CH 1
Fast–rise step
generator
50 W termination
Output
Digitizing oscilloscope
Figure 5–3: Attenuator adjustment setup and locations

Table of Contents

Other manuals for Tektronix TDS 360

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 360 and is the answer not in the manual?

Tektronix TDS 360 Specifications

General IconGeneral
BrandTektronix
ModelTDS 360
CategoryTest Equipment
LanguageEnglish

Related product manuals