Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–61
<Command> is the command that caused the error and may be returned when a
command error is detected by the digitizing oscilloscope. As much of the
command will be returned as possible without exceeding the 60 character limit
of the <Message> and <Command> strings combined. The command string is
right-justified.
ALLEV?
might return the string :ALLEV 2225,"Measurement error, No waveform to
measure; ",420,"Query UNTERMINATED; ".
ALLOcate? (Query Only)
Returns the number of data points allocated for all four reference memory
locations.
Save and Recall
ALLOcate?
ALLOcate ?
ALLOCATE?
might return :ALLOCATE:WAVEFORM:REF1 50000;REF2 0;REF3 0; REF4 0;,
indicating that all 50000 data points are allocated to reference memory loca-
tion 1.
ALLOcate:WAVEform? (Query Only)
Returns the number of data points allocated for all four reference memory
locations.
Save and Recall
ALLOcate:WAVEform?
WAVEform ?ALLOcate :
ALLOCATE?
might return :ALLOCATE:WAVEFORM:REF1 500;REF2 500;REF3 500; REF4 0;,
Examples
Group
Syntax
Examples
Group
Syntax
Examples

Table of Contents

Other manuals for Tektronix TDS 500C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 500C and is the answer not in the manual?

Tektronix TDS 500C Specifications

General IconGeneral
BrandTektronix
ModelTDS 500C
CategoryTest Equipment
LanguageEnglish

Related product manuals