Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–95
*DDT?
<QString>
*DDT
<Space>
?
<Block>
<Block> or <QString> is a complete sequence of program messages. The
messages can contain only valid commands that must be separated by semico-
lons and must follow all rules for concatenating commands (see page 2–4). The
sequence must be 80 characters. <Block> format is always returned as a query
response.
*DDT #0ACQUIRE:STATE RUN;BELL<EOI>
specifies that the acquisition system will be started and the bell rings each time a
*TRG command is sent.
DELEte:SETUp (No Query Form)
Removes stored setups from memory and initializes the location with the factory
default setup.
NOTE. The setup information cannot be recovered once it has been deleted.
Save and Recall
*RCL, RECAll:SETUp, *RST, *SAV, SAVe:SETUp, TEKSecure
DELEte:SETUp { <NR1> | ALL }
DELEte :
SETUp
<Space>
ALL
<NR1>
<NR1> is a value in the range 1 to 10, and specifies a setup storage location.
Using an out-of-range value causes an execution error.
ALL specifies all the stored setups.
Arguments
Examples
Group
Related Commands
Syntax
Arguments

Table of Contents

Other manuals for Tektronix TDS 520B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 520B and is the answer not in the manual?

Tektronix TDS 520B Specifications

General IconGeneral
BrandTektronix
ModelTDS 520B
CategoryTest Equipment
LanguageEnglish

Related product manuals