Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–97
Status and Error
*CLS, *ESE, *ESR?, EVENT?, EVMsg?, *SRE, *STB?
 


 
 is a value in the range from 0 to 255. The binary bits of the DESER are
set according to this value. For example,   sets the DESER to the binary
value 11010001 (that is, the most significant bit in the register is set to 1, the
next most significant bit to 1, the next bit to 0, etc.).
The power-on default for DESER is all bits set if  is 1. If  is 0, the
DESER maintains its value through a power cycle.
NOTE. Setting the DESER and the ESER to the same value allows only those
codes to be entered into the Event Queue and summarized on the ESB bit (bit 5)
of the Status Byte Register. Use the  command to set the ESER. A discussion
of event handling begins on page 3–1.
 
sets the DESER to binary 11010001, which enables the PON, URQ, EXE, and
OPC bits.

might return the string  , showing that the DESER contains the binary
value 10111010.
DIAg:RESUlt:FLAg? (Query Only)
Returns the pass/fail status from the last diagnostic test sequence execution. The
DIAg:RESUlt:LOG? query can be used to determine which test(s) has failed.
Calibration and Diagnostic
Group
Related Commands
Syntax
Arguments
Examples
Group

Table of Contents

Other manuals for Tektronix TDS 520B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 520B and is the answer not in the manual?

Tektronix TDS 520B Specifications

General IconGeneral
BrandTektronix
ModelTDS 520B
CategoryTest Equipment
LanguageEnglish

Related product manuals